LFY-716 Exhalation Valve Air Tightness Tester
Main purpose
The airtightness test of the exhalation valve for self-suction filtering particulate respirators. Complies with the test standard GB 2626-2006 "Respiratory Protective Equipment - Self-Suction Filtering Anti-Particulate Respirators". Applicable to labor safety protection inspection centers, occupational safety inspection centers, disease prevention and control centers, respirator manufacturing enterprises, etc.
Technical indicators
1. Air source: Vacuum pump;
2. Constant volume chamber: Volume (150 ± 10) mL;
3. Gas flowmeter: Range (0 - 1000) mL/min;
4. Manometer: Measuring range: ±1000 Pa;
5. Timer: 0 - 99.99 seconds per minute/hour, adjustable as needed;
6. Power Supply: 220V, 50Hz.
7. The vacuum pump's air suction rate is 2 liters per minute.
Applicable Standards
GB 2626-2006 Respiratory Protective Equipment - Self-Contained Filtering Respirator for Particulate Matter
Contact Us
Address: No. 195, Shandong Road, Shibei District, Qingdao City, Shandong Province
E-mail: tzfzyq@163.com
Contact Number
Instrumentation and Equipment: 0532-85614755 0532-85635588 13475428416
Inspection and Testing: 0532-85623450 0532-85634678
Scientific Journals: 0532-85648088
Scientific Services: 0532-85641981
Video Channel