LFY-602A Large 45° Fabric Flame Resistance Testing Instrument
Main purpose
Method A is used to measure the damaged area and length of textiles and other materials when they are burned at a 45-degree angle. Method B is used to determine the number of times textiles come into contact with flames when they are subjected to thermal melting up to a specified length at a 45-degree angle.
Technical indicators
Reignition timing range: 0 - 99.99 seconds; Resolution: 0.01 seconds
Ignition time range: 0 - 99.99 seconds; Resolution: 0.01 seconds Distance from the ignition device to the sample
Ignition time: Method A: Thin fabric 30 seconds; Thick fabric 120 seconds Method B: Manual
Ignition device diameter: Method A: Thin fabric Φ6.4mm; Thick fabric Φ20mm Method B: Φ6.4mm
Sample holder size: Method A: Overall size 500mm × 230mm; Inner frame 250mm × 150mm
Method B: Inner diameter of the coil is 10mm; Pitch is 2mm; Total length is 150mm
Distance from the top of the igniter to the sample: Method A: 45mm for thin fabric; 65mm for thick fabric; Method B: 45mm
0 to 99.99 seconds; Resolution: 0.01 seconds
Applicable standard
GB/T 14645-2014 Textiles - Determination of the damaged area and the number of flame extensions in the 45° direction for burning performance
GB/T 17591-2025 Flame Retardant Fabrics
Contact Us
Address: No. 195, Shandong Road, Shibei District, Qingdao City, Shandong Province
E-mail: tzfzyq@163.com
Contact Number
Instrumentation and Equipment: 0532-85614755 0532-85635588 13475428416
Inspection and Testing: 0532-85623450 0532-85634678
Scientific Journals: 0532-85648088
Scientific Services: 0532-85641981