LFY-604D Flame Spread Testing Instrument
Main purpose/Use
The Flame Spread Tester is a device that complies with ISO15025-2016 and is used to measure the limited flame spread performance of industrial products in the form of directional textile fabrics and single-component or multi-component fabrics (coatings, quilting, multi-layer, sandwich structures and similar combinations) under a small-scale flame. This test method is not applicable to materials that exhibit extensive melting or shrinking.
Apply the pre-determined flame to the surface (surface ignition) or the bottom (edge ignition) of the vertically suspended sample. The application time is 10 seconds. Measure and observe the rate of flame propagation, the form of the dripping substances, whether the burning dripping substances form holes, afterglow, and the duration of re-ignition, and measure the length of the damage. Based on these, evaluate the flame retardant performance of the fabric.
Technical Specifications
Sample size: Bottom material: 230mm × 230mm (±3mm) Four panels in both warp and weft directions; Top material: 90mm × 300mm (±3mm) Four panels in both warp and weft directions.
Flame height during combustion: 40±2mm (height adjustable)
Ignition time: 12s (preset adjustable)
Reignition time: 1s to 999.9s (digital display, manual pause allowed for display)
Ignition nozzle angle: Vertical method
Ignition nozzle diameter: 5mm inner diameter gun tube with a fixed throttle hole 75mm ± 6mm away
Burner: Width 320mm × Depth 320mm × Height 760mm
Combustion gas: Propane gas purity should be above 95%
Inlet air pressure of igniter: 2.5psi ± 0.25Psi (adjustable)
Test power supply: 220V 0.5kVA
Lighting pipe diameter: Approximately 1.6mm, 3mm away from the edge of the burner, flame length of the light source 3mm
Cabinet: Provides an independent space for the test
Observation window: Transparent tempered glass observation window for convenient observation of the test process
Fixing block for sample: Fixed the sample holder to keep it vertical and in the middle position
Sample holder: Sample fixing fixture
Flame measuring ruler: Flame height measuring ruler
Ventilation opening: Provides necessary air circulation for the combustion test
Environmental requirements: Temperature (23±5°C) humidity between 15% and 80%
Control method: Ordinary electromechanical control, easy to operate
Overall dimensions: Width 1120mm × Depth 580mm × Height 1350mm
Applicable standards: ISO 15025:2016 Protective clothing - Fire protection - Test method for limited flame spread ISO 11612:2015 Protective clothing - Heat insulation and flame retardant protective clothing - Minimum performance requirements
Installation Requirements
1. It is recommended that the equipment be placed on a workbench that is at least 70 cm high, with the tabletop size being larger than 120 cm by 65 cm.
2. Open an exhaust vent with a diameter of more than 10 cm at a height of more than 140 cm from the center of the countertop.
3. The equipment itself has no specific requirements for the environment. All that is needed is to adjust the temperature and humidity to the specified range when conducting each standard test.
Contact Us
Address: No. 195, Shandong Road, Shibei District, Qingdao City, Shandong Province
E-mail: tzfzyq@163.com
Contact Number
Instrumentation and Equipment: 0532-85614755 0532-85635588 13475428416
Inspection and Testing: 0532-85623450 0532-85634678
Scientific Journals: 0532-85648088
Scientific Services: 0532-85641981
Video Channel